Vue de la page 0
Probe head with
Diaphragm set
Probe filling Electrolyte
SPECIFICATIONS
Ranges: Dissolved Oxygen 0 to 20.0 mg/L (liter)
Oxygen hi Air 0 to 100.0% 0
2
Temperature 0 to 50°C
Resolution: Dissolved Oxygen 0.1 mg/L
Oxygen 0.1% 0
2
Temperature 0.1°C
Accuracy: Dissolved Oxygen ± 0.4 mg/L
Oxygen ±0.7% 0
2
Temperature ±0.8°C, ±1.5°F
Probe Compensation Temp: 0 to 50 °C, automatic
and Adjustment: Salt: 0 to 39% Salt
Height 0 to 3900 meters
Sampling Time: approximately 0.4 seconds
Data Output: RS-232 PC serial Interface
Case: ABS plastic
Fail safe: Low battery Indicator
Power: one (1) 9-Volt alkaline battery
Accessories
Supplied: carrying case, oxygen probe, spare
diaphragms (two), electrolyte for probe,
battery, Traceable
®
Certifi cate, instructions.
FIGURE 1
DESCRIPTION
4-1 LCD Display
4-2 Power on/ff Button
4-3 Data Hold Button: press to hold reading
4-4 °C/°F button
4-5 LCD contrast Adjustment Knob
4-6 Memory Record Button: press to record results.
4-7 Memory Recall Button: press to show minimum and
maximum readings.
4-8 Factor Adjustment Button
4-9 Zero Button
4-10 % Salt Button
4-11 DO/0
2
Button
4-12 0
2
Calibration Button
4.43 Mountain Height Button
4-14 Battery Compartment
4-15 Probe Input Socket
4-16 RS-232 Output Socket
4-17 Oxygen Probe
4-18 Temperature sensor
4-19 Probe head with diaphragm
4-20 Protection cover for probe head
4-21 Oxygen Probe Plug
DIGITAL OXYGEN METER OPERATION
This unit has been shipped with Electrolyte Liquid 1.
in the sensor. This enables the probe to correctly
report oxygen. If calibrating the meter is diffi cult or if
the meter readings are unstable, then it is possible
that there is insuffi cient or no electrolyte in the
sensor. SEE REPLACING ELECTROLYTE on how
to fi ll the sensor.
Before use, the unit must be calibrated. Disconnect 2.
the OXYGEN PROBE PLUG (4-18, fi g. 1) from the
socket on top of the unit labeled INPUT (4-15, g. 1).
Turn the meter on by switching the POWER 3.
BUTTON (4-2, fi g. 1) to the right.
Select 04.
2
by sliding the O
2
/DO selector (4-11, fi g. 1)
to this position.
Press the ZERO BUTTON (4-9, fi g. 1). The display 5.
will show 0.
Connect the OXYGEN PROBE PLUG (4-18, 6.
g. 1) to the socket on top of the unit labeled INPUT
(4-15, fi g. 1). Wait at least fi ve (5) minutes until the
displayed values stabilize and no longer fl uctuate.
Press the 07.
2
CALIBRATION BUTTON (4-12, fi g. 1).
The display will show either 20.8 or 20.9, the typical
oxygen percentage in the air.
NOTE: Calibrate the meter In a large, well
ventilated environment for best results.
To measure Dissolved oxygen, slide the O8.
2
/DO
selector (4-11, fi g. 1) to the DO position.
Depending on your application, you may need to 9.
adjust the % salt compensation of the probe. Press
% SALT BUTTON (4-10, fi g. 1). The Display will
show an “S” for salinity, and 0%. Press the FACTOR
ADJUSTMENT BUTTON (4-8, fi g. 1) once. This will
add 1% to the original salt %. Continue pressing
this button until it reaches the desired value. When
complete, press % SALT BUTTON (4-10, fi g. 1).
If your measurement is not taking place at sea level, 10.
you will need to adjust the Height compensation.
Press the MT BUTTON (4-13, fi g. 1) The Display will
show an “H” for height and a 0 for zero sea level.
Press the FACTOR ADJUSTMENT BUTTON
(4-8, fi g. 1) once. This will add 100 meters. Continue
pressing this button until it reaches the desired
value. When complete, press MT BUTTON
(4-13, fi g. 1).
Immerse the probe at least 10 cm into the liquid 11.
being measured. This ensures that the probe will
measure the temperature of the liquid and the
automatic temperature adjustment will take place.
Allow a few minutes for the probe temperature reach
the temperature of the liquid. If there is more than a
few degrees of difference between the temperature
of the liquid and the probe, allow more time for the
probe temperature to adapt.
To measure the dissolved oxygen content, the 12.
velocity of the liquid being measured must be at
least 0.2-0.3 m/s. To achieve this, immerse the
probe in the solution and gently shake it. For a more
accurate measurement, use a magnetic agitator to
ensure a certain velocity. Indicate that the value is a
“held” value. To cancel the data hold feature, simply
press the HOLD BUTTON a second time.
After use, rinse the probe thoroughly with tap water.13.
DATA HOLD
To hold a measurement on the display, press the HOLD
BUTTON (4-3, fi g. 1) while a measurement is being
taken. The LCD display will show DH in the upper left
portion of the display to indicate that the value is a “held”
value. To cancel the data hold feature, simply press the
HOLD BUTTON a second time.
DATA RECORD
To record a measurement, press the RECORD 1.
BUTTON (4-6, fi g. 1) while a measurement is being
taken to hold the displayed value. The LCD display
will show REC in the lower left portion of the display
to indicate that the value is being recorded. To
deactivate the record function, press the RECORD
BUTTON again.
Memory Recall: Use the RECALL BUTTON 2.
(4-7, fi g. 1) to recall the minimum, maximum and
average readings. When the REC symbol appears on
the display, press the RECALL BUTTON (3-6, fi g. 1)
once. The maximum recorded value will be displayed.
The letters “Max” will also appear on indicating that
this is the maximum or highest reading. A second
press of the RECALL BUTTON will display the
minimum reading. The letters “Min” will also appear
indicating that this is the minimum or lowest reading.
A third press of the RECALL BUTTON will display the
average reading. The letters “Avg” will also appear
indicating that this is the average reading. Note: The
Data Record function must be in use to utilize the
Memory recall features. Once the RECORD BUTTON
has been pressed a second time to deactivate the
data record function, the minimum, maximum and
values are no longer stored.
AUTOMATIC SHUT-OFF
The unit has an automatic shut off feature to prolong
battery life. If no button on the unit is pressed for ten
minutes, the unit will automatically shut off. To deactivate
this feature, press the RECORD BUTTON while a
measurement is being taken.
DISPLAY CONTRAST ADJUSTMENT
The Contrast of the LCD display may need to be
changed either because of light conditions, battery
voltage drop, or user preferences. The LCD display
contrast can be adjusted using the LCD DISPLAY
CONTRAST ADJUSTMENT KNOB (4-5, fi g. 1). Simply
turn the knob to the left or right until the optimal LCD
contrast is obtained. If the display is fi lled with “88888”
when no reading is being taken, the contrast needs to be
adjusted. Turn the contrast knob to the left until a single
“0” appears.
RESETTING THE CPU SYSTEM
Certain incorrect operational procedures may disrupt the
meters CPU system. If the meter is not working properly,
the circuit may need to be reset. Set the DO/0
2
switch to
0
2
(4-11, fi g. 1). Turn off the meter by pressing the power
button (4-2, fi g. 1). Disconnect the probe from the meter.
Press the 0
2
CALIBRATION BUTTON (4-12, fi g. 1)
continuously. While still pressing this button, press the
power button again. Then press the zero button (4-9,
g. 1). Reconnect the probe and wait until the reading
stabilizes. Press the 0
2
CALIBRATION BUTTON (4-12,
g. 1) again. Turn the power off. Begin normal operating
procedures again.
FIGURE 2
REPLACING THE ELECTROLYTE
When the meter cannot be calibrated 1.
properly or if the reading is unstable, the
electrolyte may need to be refi lled or the
diaphragm may be dirty and need to be
replaced. Unscrew the Probe Head
(8-3, fi g. 2)
Pour out the old electrolyte from the Probe Head 2.
(8-3, fi g. 2).
Pour in the new Electrolyte into the container of the 3.
probe head.
Reassemble the Probe Head (8-3, fi g. 2) with the Probe 4.
Body (8-1, fi g. 2).
5.
Vue de la page 0
1 2

Résumé du contenu

Page 1 - Probe filling Electrolyte

Probe head with Diaphragm setProbe filling ElectrolyteSPECIFICATIONSRanges: Dissolved Oxygen 0 to 20.0 mg/L (liter) Oxygen hi Air 0 to 100.0% 02

Page 2 - INSTRUCTIONS

©2007 Control Company. 92-4320-00 Rev. 0 103007RS—232 PC SERIAL INTERFACEThis unit features RS—232 output. A RS—232 OUTPUT SOCKET (4-10, fi g. 1) is

Commentaires sur ces manuels

Pas de commentaire